Specifications

When you need precise, rapid results from an extremely small sample, the only answer is the Fiske 210 Micro-Sample Osmometer.
The Fiske 210 continues our 50-year commitment to osmometry - a commitment that is well known to thousands of users around
the world.
The Fiske 210 freezing-point osmometer requires only a 20μL sample size, which provides accurate osmolality results in less than 90 seconds. For convenience, the fully accessible probe design is easy to clean and service. The Fiske 210 is not only your most reliable option, but also the easiest to use. Just pipet 20μL of sample into a disposable sample tube, then place the tube into the test chamber. Lower the probe into the sample and push the “Test” button. Results are displayed in ninety seconds.
Microprocessor control ensures repeatability and accuracy. The Fiske 210 calibrates automatically at the push of a button, with no user adjustments. Because calibration is highly stable, however, the procedure is seldom required. Self-diagnosticsalert the operator to any malfunctions or abnormal test conditions through messages on the display.
Since the Fiske 210 Osmometer doesn't use a liquid cooling bath, you'll have minimal routine maintenance. In fact, you don't have to worry about preventive maintenance, either. All these features save you costly operator labor and downtime.

 

CLINICAL APPLICATIONS OF OSMOMETRY:Determine Osmol Gap – actual versus calculated osmolality is a rapid indicator for toxicology and trauma cases. Remember, only the freezingpoint method is sensitive to alcohol and ketones.
Hydration Status for Athletes – weight class and endurance events are conducive to dehydration, with performance and health consequences.
Mannitol Treatment – maintain, without exceeding, critical level of therapeutic infusion.
Glycine Ingestion – monitor the absorption of irrigation fluids in surgical procedures.
Renal Function – evaluate concentrating ability and monitor ADH therapy.
Differentiate DKA and NKH – only a measured osmolality can make this determination and guide effective therapy.
Sample Authentication for Drug Screens – low urine osmolality is indicative of tampering.

 
A:Alphanumeric display
Results, user prompts, and instrument diagnostics are clearly displayed.
B:Easy push-button operation
Requires minimum operator training.
C:Printer, RS232C and barcode scanner ports
A full range of data handling options are available. Unit connects easily to printer, computer, and scanner.
D:Automatic calibration
No fine-tuning or user adjustments during calibration. The instrument’s microprocessor automatically makes calibration adjustments at the push of a button.
E:Smallest sample size
Only 20 microliters of sample needed.
F:Solid-state cooling
The 210 has no liquid cooling bath; no maintenance required.

Model 210 Single-Sample Osmometer Specifications:
Sample Volume 20 μL
Test Time 90 seconds
Sample Capacity Single Sample
Units mOsm/kg H2O
Resolution 1 mOsm/kg H2O
Range 0 to 2000 mOsm/kg H2O
Communications DTE RS-232 serial port, parallel printer port, and optional barcode scanner
Performance at Reference Conditions1
Linearity Less than ±1% from a straight line between 0 and 2000 mOsm/kg H2O
Repeatability ±2 mOsm/kg H2O (1 S.D.) between 0 and 400 mOsm/kg H2O;
±0.5% (1 S.D.) between 400 and 2000 mOsm/kg H2O
Drift Less than 1 mOsm/kg H2O per month
Performance Over Operating Conditions
Temperature Effects Less than 1 mOsm/kg H2O per 5°C (9°F) ambient temperature change
Operating Conditions
Temperature 18 to 35°C (64 to 95°F)
Humidity 5 to 80% relative humidity, (non-condensing)
Storage Temperature -40 to +45°C (-40 to +113°F)
Electrical Voltage 100 to 240 VAC (50/60 Hz)
Power Consumption 60 Watts
Dimensions 12.0" H x 12.0" W x 9.5" D
(30.5 cm x 30.5 cm x 24.1 cm)
Net Weight 13.6 lbs. (6.2 kg)
Shipping Weight 20.5 lbs. (9.3 kg)
Warranty One-year limited warranty on workmanship and all parts except glass, plastic, and parts warranted by their makers
1Reference Conditions 1Reference Conditions
Specifications subject to change

In order to obtain peak and long-lasting performance from your osmometer, ask about these related products:
Accuref™ 290 Reference routinely checks calibration stability.
Protinol\® Near-Serum Reference, in 3 levels, gives added confidence in interlab surveys.
Renol™ is the world’s only osmometer-specific control solution for laboratories testing urine samples.
Linearity Set™ documents performance, per CLIA guidelines, over full instrument range.
Fiske® Calibration Standards and Sample Kits are manufactured to exacting tolerances to ensure that your osmometer operates at peak performance.